张帆,王江,池海江,等. 怀4井ATG-6138M气汞仪干扰因素及质量评价[J]. 华北地震科学,2021, 39(2):80-86. doi:10.3969/j.issn.1003−1375.2021.02.012.
引用本文: 张帆,王江,池海江,等. 怀4井ATG-6138M气汞仪干扰因素及质量评价[J]. 华北地震科学,2021, 39(2):80-86. doi:10.3969/j.issn.1003−1375.2021.02.012.
ZHANG Fan,WANG jiang,CHI Haijiang,et al. Interference Factors and Quality Evaluation of ATG-6138M Vapormeter in Huailai No.4 Well [J]. North China Earthquake Sciences,2021, 39(2):80-86. doi:10.3969/j.issn.1003−1375.2021.02.012.
Citation: ZHANG Fan,WANG jiang,CHI Haijiang,et al. Interference Factors and Quality Evaluation of ATG-6138M Vapormeter in Huailai No.4 Well [J]. North China Earthquake Sciences,2021, 39(2):80-86. doi:10.3969/j.issn.1003−1375.2021.02.012.

怀4井ATG-6138M气汞仪干扰因素及质量评价

Interference Factors and Quality Evaluation of ATG-6138M Vapormeter in Huailai No.4 Well

  • 摘要: 以怀来后郝窑地震台怀4井ATG-6138M型气汞仪为研究对象,分析及排除其观测干扰因素,为正确识别震前异常信息提供参考依据;对ATG-6138M型气汞仪和DFG-B型气汞仪进行观测质量对比评价,运用origin数据统计软件进行数据分析。研究表明:ATG-6138M型气汞仪数据变化较为灵敏,且观测质量优于DFG-B型气汞仪,同时发现怀4井气汞浓度主要受到观测室温度和动水位的影响。

     

    Abstract: Taking the ATG-6138M vapormeter of Huailai No.4 Well in Houhaoyao seismic station as the research object, we analyzed and excluded the interference factors of its observation, and provided a reference for correctly identifying the pre-earthquake anomaly information. The observation quality of ATG-6138M vapormeter and the DFG-B vapormeter are contrastively analyzed using origin data statistical software. The result shows that: data changes of the ATG-6138M vapormeter are more sensitive and its observation quality is better than that of the DFG-B vapormeter. At the same time, it is found that the concentration of gas mercury in Huailai No.4Well is mainly affected by the temperature of observation room and dynamic level in well.

     

/